ASME V Article 2 · Ug screen
3 in object at the exact Ug limit
A 0.030 in effective X-ray focus images a 3 in object at 6 in SFD with the detector in contact.
Reading the result. The source-to-object distance equals the object thickness, so Ug equals the effective focus. Equality passes this geometry-only screen.
Setup
Exposure geometry
mm
mm
Film-side surface in contact.
mm
Reference shot
mm
min
GBq
Activity at the reference shot.
GBq
d
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Calculated Ug
- 0.762mm
- ASME Ug limit
- 0.762mm
- Minimum SFD
- 152.4mm
- Exposure time
- 0.28674min
Pass · within Ug limit
This verdict covers the geometric unsharpness screen only.
- Source-to-object distance
- 76.2mm
- Ug margin to limit
- 0mm
- Inverse-square time factor
- 0.1147×
- Source half-life
- —
- Decay-adjusted activity
- —
- Remaining activity
- —
- Activity correction
- 1×
Use the project-specified Section V edition and its native limit table. The exposure estimate assumes the same object, detector and beam conditions as the reference. Verify the approved technique and image-quality indicators. Equations and references