ASME V Article 4 · ISO 10863
UT TOFD Weld Sizing & Dead Zone
Size a flaw from two longitudinal tip arrivals, check probe center separation and estimate the lateral-wave obscured zone.
Setup
Probe geometry
mm
mm
°
PCS target places crossing at ⅔ thickness.
m/s
Arrival time picks
mm
m/s
µs
µs
µs
µs
0 uses one sample as a lower bound.
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Top tip depth
- 9.7396mm
- Bot. tip depth
- 21.572mm
- Flaw height
- 11.833mm
- Lateral zone
- 8.156mm
Tip picks clear the timing screen
Depths assume the two tips lie on the probe-pair midline.
- PCS target at ⅔ wall
- 69.282mm
- Lateral arrival in chosen datum
- 15.126µs
- Lateral material travel time
- 11.703µs
- Backwall arrival in chosen datum
- 18.905µs
- Backwall material travel time
- 15.482µs
- Two-wedge delay
- 3.4232µs
- Tip-pick separation
- 1.63µs
- Modeled obscuring interval
- 0.32µs
- Depth change per sample at top tip
- 0.2165mm
Use a calibrated material velocity, actual probe index separation and measured pulse duration. Off-midline reflectors, wedges and wave overlap require a qualified TOFD setup. Equations and references
Worked examples
Open a setup to inspect its inputs, then try your own values.