ASME V Article 4 · ISO 10863

Top pick coincident with the lateral wave

The upper signal is at exactly zero lag from the lateral wave, while the lower pick is 0.52 µs later.

Reading the result. The top calculation gives depth zero. Timing overlap flags the setup because the tip is inside the lateral obscured zone.

Setup

Probe geometry

mm
mm
°
PCS target places crossing at ⅔ thickness.
m/s

Arrival time picks

µs
µs
µs
µs
0 uses one sample as a lower bound.
Worked examples

Results update as you edit. Use calibrated measurements.

Inspection readout

SI units
Top tip depth
0mm
Bot. tip depth
9.7997mm
Flaw height
9.7997mm
Lateral zone
4.2532mm
Top tip in lateral obscured zone

The top pick lies within the modeled pulse or sample window. Resolve the indication on a calibration block.

Probe pair and TOFD arrivalsThe cross-section shows the midline flaw model. The timing sketch traces lateral, tip and backwall arrivals as a scan passes the flaw.Pitch-catch setupTime tracesLateral → backwall0.52 µs apartLateralTip pathsBackwall
The cross-section shows the midline flaw model. The timing sketch traces lateral, tip and backwall arrivals as a scan passes the flaw.
PCS target at ⅔ wall
69.282mm
Lateral arrival in chosen datum
0µs
Lateral material travel time
10µs
Backwall arrival in chosen datum
4.1421µs
Backwall material travel time
14.142µs
Two-wedge delay
—
Tip-pick separation
0.52µs
Modeled obscuring interval
0.1µs
Depth change per sample at top tip
1.8983mm

Use a calibrated material velocity, actual probe index separation and measured pulse duration. Off-midline reflectors, wedges and wave overlap require a qualified TOFD setup. Equations and references