ASME V Article 4 · ISO 10863

30 mm weld with two-wedge delay

A 60° probe pair on 30 mm steel has an instrument clock referenced before the wedges. The tip arrivals are booked at 15.58 and 17.21 µs.

Reading the result. Removing the two-wedge delay places the top and bottom at about 9.74 and 21.57 mm. The flaw height is about 11.83 mm.

Setup

Probe geometry

mm
mm
°
PCS target places crossing at ⅔ thickness.
m/s

Arrival time picks

mm
m/s
µs
µs
µs
µs
0 uses one sample as a lower bound.
Worked examples

Results update as you edit. Use calibrated measurements.

Inspection readout

SI units
Top tip depth
9.7396mm
Bot. tip depth
21.572mm
Flaw height
11.833mm
Lateral zone
8.156mm
Tip picks clear the timing screen

Depths assume the two tips lie on the probe-pair midline.

Probe pair and TOFD arrivalsThe cross-section shows the midline flaw model. The timing sketch traces lateral, tip and backwall arrivals as a scan passes the flaw.Pitch-catch setupTime tracesLateral → backwall1.63 µs apartLateralTip pathsBackwall
The cross-section shows the midline flaw model. The timing sketch traces lateral, tip and backwall arrivals as a scan passes the flaw.
PCS target at ⅔ wall
69.282mm
Lateral arrival in chosen datum
15.126µs
Lateral material travel time
11.703µs
Backwall arrival in chosen datum
18.905µs
Backwall material travel time
15.482µs
Two-wedge delay
3.4232µs
Tip-pick separation
1.63µs
Modeled obscuring interval
0.32µs
Depth change per sample at top tip
0.2165mm

Use a calibrated material velocity, actual probe index separation and measured pulse duration. Off-midline reflectors, wedges and wave overlap require a qualified TOFD setup. Equations and references