Gas bleed & SCCM conversion / Worked example
CO₂ Throughput From Torr·L/s
A CO₂ dosing stage reports 0.045 Torr·L/s at a 0 °C reference and runs at 315 K in the chamber.
Inputs
Live readout
Flow converted- SCCM
- 3.553 SCCM
- SLM
- 3.553e-3 SLM
- Torr·L/s
- 0.045 Torr·L/s
- mbar·L/s
- 0.06 mbar·L/s
- Pa·m³/s
- 6e-3 Pa·m³/s
- mol/s
- 2.642e-6 mol/s
- mg/s
- 0.1163 mg/s
- Throughput at chamber temperature
- 6.919e-3 Pa·m³/s
- Predicted process pressure
- 0.7687 Pa
- Downstream / upstream ratio
- Known flow mode
- Critical pressure ratio
- Known flow mode
- Orifice upstream Kn
- Known flow mode
Reading this example
Use the warmer chamber throughput for process pressure. It differs from the throughput at the standard-temperature reference.
Model & assumptions
Reference flow
ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM
SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.
Orifice choking
r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁
Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.