Gas bleed & SCCM conversion / Worked example
Oxygen Mass Flow From mg/s
An oxygen source specifies 0.8 mg/s. Back out the standard volume flow and chamber pressure at 24 L/s.
Inputs
Live readout
Flow converted- SCCM
- 33.62 SCCM
- SLM
- 0.03362 SLM
- Torr·L/s
- 0.4259 Torr·L/s
- mbar·L/s
- 0.5678 mbar·L/s
- Pa·m³/s
- 0.05678 Pa·m³/s
- mol/s
- 2.5e-5 mol/s
- mg/s
- 0.8 mg/s
- Throughput at chamber temperature
- 0.06094 Pa·m³/s
- Predicted process pressure
- 2.539 Pa
- Downstream / upstream ratio
- Known flow mode
- Critical pressure ratio
- Known flow mode
- Orifice upstream Kn
- Known flow mode
Reading this example
Molar mass is essential when starting from mass flow. The same mg/s for another gas does not give the same SCCM.
Model & assumptions
Reference flow
ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM
SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.
Orifice choking
r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁
Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.