Gas bleed & SCCM conversion / Worked example

Oxygen Mass Flow From mg/s

An oxygen source specifies 0.8 mg/s. Back out the standard volume flow and chamber pressure at 24 L/s.

Readout ↓

Inputs

Flow basis and process chamber
°C
L/s

Live readout

Flow converted
Standard flow33.62SCCMMass flow0.8mg/sChamber load0.06094Pa·m³/sProcess P2.539PaP₁Gas supplyQMFCP₂Process2.54 PaProcess pressure · Pa vs SCCM02.094.19027.755.5Reference0 °C / 1 atmP₂ / P₁Known flowCritical ratio—
SCCM
33.62 SCCM
SLM
0.03362 SLM
Torr·L/s
0.4259 Torr·L/s
mbar·L/s
0.5678 mbar·L/s
Pa·m³/s
0.05678 Pa·m³/s
mol/s
2.5e-5 mol/s
mg/s
0.8 mg/s
Throughput at chamber temperature
0.06094 Pa·m³/s
Predicted process pressure
2.539 Pa
Downstream / upstream ratio
Known flow mode
Critical pressure ratio
Known flow mode
Orifice upstream Kn
Known flow mode

    Reading this example

    Molar mass is essential when starting from mass flow. The same mg/s for another gas does not give the same SCCM.

    Model & assumptions

    Reference flow

    ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM

    SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.

    Orifice choking

    r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁

    Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.